Diagnostic: Gradient step-test

This is a variation on the dwell volume measurement. Instead of running a linear gradient, "stairstep" the composition (e.g., 0% for 5 minutes, 10% for 5 minutes, 20% for 5 minutes, etc.). 
  • If  the absorbance shift at each step is "identical" (i.e., the variations are within the manufacturer's specification), then the proportioning system is functioning correctly. 
     
  • If the variation is absorbance is greater than that specified for the proportioning system, the system should be serviced. Proportioning problems can result from sticking or defective switching valves, sticking, dirty, or defective check valves, air bubbles in check valves or pump heads, or software errors. 
     